The Department of Veterans Affairs is seeking quotations for sterilization containers. The solicitation includes a list of salient characteristics that the containers must meet, such as compatibility with various sterilization cycles, ergonomic handles, and filter retention plates. The containers must also be validated for stacking during sterilization and storage. The solicitation is an "or equal" solicitation, meaning that the brands shown are examples, and other brands may be considered if they meet the requirements. The contractor must provide manufacturer's instructions for use and ensure that the containers are compatible with automated washer disinfectors.
• The containers must be compatible with Prevac, IUSS, Gravity, Steam, and low temp sterilization for STERRAD 100NX machines.
• The containers must have ergonomic handles and filter retention plates with audible click or valve to replace filter.
• The containers must be validated for stacking during sterilization and storage.
• The contractor must provide manufacturer's instructions for use.
• The containers must be compatible with automated washer disinfectors with ability to withstand temperatures of 190 degrees.