The Argonne National Laboratory is seeking a laboratory steam sterilizer with on-site installation and removal of the old unit. The sterilizer should have a single door, pass-through configuration, and an internal volume of 9.7 cu. ft. The laboratory is looking for a firm fixed price quotation to supply the materials and services as indicated. The contractor's work on the Argonne site is subject to compliance with the ES&H requirements. The contractor is responsible for the removal and off-site disposal of any debris resulting from the performance of this work.
Some key features of the required sterilizer include:
• Internal Volume: 9.7 cu. ft. (275 L)
• Door Configurations: Single Door, Pass-through
• Design Operating Temperature: Gravity/Vacuum Cycles: 230–275°F (110–135°C), Liquid Cycles: 219–275°F (104–135°C), Optional Low Temp Cycle: 169–212°F (76–100°C)
• Interior Dimensions (w × h × d): 21" × 21" × 38" (532 × 532 × 965 mm)
• Exterior Dimensions (w × h × d): Single Door: 30" × 74" × 68.125" (762 × 1880 × 1222 mm)
• Door Operation: manual (standard)
• Steam Source: facility/house steam (standard)
• Air Removal: ejector (standard)
• Recesses and Barrlers: freestanding (cabinet enclosure panels required)