The Department of the Army is seeking a contractor to provide exam lights recessed for the Evans Army Community Hospital. The lighting system must be operable without compromising the sterile field required for vaginal deliveries and provide bright, focused illumination deep into the birth canal. The system must also be compatible with standard hospital ceiling heights and grid systems, operate on standard hospital electrical systems, and include both a wall-mounted control panel and a handheld wand. The contractor will be responsible for installation, inspection, and warranty certification of the recessed lighting system.
Some key features of the required system include:
• The light must provide a minimum level of brightness at a standard working distance (14,000 LUX at 3 meters and 49,000 LUX at 1 meter).
• A single controller must be able to operate multiple light heads simultaneously.
• The light source must be energy-efficient, long-lasting, and produce minimal heat.
• The system must allow for precise, remote aiming of the light beam from a distance via remote or wand.
The contractor will also be responsible for removing existing equipment from the site and coordinating the disposition of the equipment.