The Iowa City VA Health Care System is seeking a phacoemulsification system and handpieces. The system must meet specific requirements, including dimensions, touchscreen size, frequency, and pulse mode range. The handpiece must be compatible with the Stellaris system and have a six-crystal piezoelectric design. The contractor must provide a service plan that includes annual or semi-annual preventive maintenance, unlimited on-site repair visits, and replacement of worn components.
• The system must have a minimum of 18" touchscreen and no greater than 19" touchscreen.
• The system must have a frequency of 28.5 kHz.
• The system must have a pulse mode range of 1-250 pulses per second.
• The system must have longitudinal motion type.
• The system must have gravity and/or pressurized air.
• The system must have a rotary vane vacuum pump.
• The system must have a vacuum range of 0-600 mmHg.
• The system must support 20-gauge, 23-gauge, and 25-gauge handpieces.
• The system must have a cutting rate of 30-800 cuts per minute.
• The system must have a dual-linear foot pedal (wireless or wired).
• The foot pedal must have independent control over the ultrasound, vacuum, aspiration, and irrigation.
You found this one too late. Don't find the next one too late.